R&S NRP Remote Control - Commands
1144.1400.12 6.93 E-3
SYSTem
With the aid of the SYSTem system, administrative device settings can be made and queried. This
includes:
• list of error messages
• IEC/IEEE-bus address
• resetting the R&S NRP
• setting the date and time
• audible signals
• version numbers of hardware and software
• setting the system speed
Table 6-31: Commands of the SENSe system
Command Parameter Unit Remark
SYSTem
Query only
:BEEPer
[:IMMediate] -
No query
:NOTify
:OVERload[?]
ON | OFF
-
:KEY[?] ON | OFF -
:STOP - -
No query
:TIME <NRf> S
No query
:COMMunicate
:GPIB
[:SELF]:ADDRess
0..30
No query
:DATE[?] <year>,<month>,<day>
:INFO? [<string >]
Query only
:ERRor? -
Query only
:KEY <NR1> | character_value>
No query
:PRESet -
No query
:SENSor[1..4]
:INFO?
[<string >]
Query only
:RESet -
No query
:TEST? - -
Query only
:SPEed NORMal | FAST
No query
:STANdard
:CATalog?
-
-
No query
:PRESet <string> -
No query
:PWSettings[?] ON|OFF -
:TRIGger:SOURce[?] INT|EXT -
:TIME[?] <hour>,<minute>,<second>
:VERSion? -
Query only
SYSTem:BEEPer[:IMMediate]
Generates an audible signal via the built-in loudspeaker. The duration of the signal is set with
SYST:BEEP:TIME. Frequency and volume cannot be changed. The signal is only output if it is not
switched off with SYS:BEEP:STAT OFF.
SYSTem:BEEPer:NOTify:OVERload[?] ON | OFF
Determines if the R&S NRP shall emit an acoustic signal when the power at a sensor exceeds the
power limit.
*RST-Wert: OFF