Remote Control - Fundamentals R&S NRP
1144.1400.12 5.2 E-3
Measurement Modes
To simplify operation in the manual mode, preset values are used in some cases for the trigger system
(TRIGger command) and the timing of the measurement (SENSe[1..4]:TIMing:EXCLude:STARt
and :STOP) when the measurement mode (ContAv, Burst, Timeslot, Scope)
is selected (Æ Table 5-2). This is not the case in the remote control mode. Here the trigger system and
the timing can be configured by the user as required.
Calculate Blocks
The R&S NRP manages 8 internal calculate blocks. Each calculate block has two input channels
(primary and secondary channel) to which one of up to 4 power sensors can be assigned. The channels
can be processed together by means of a selectable function. The result is available at the output of the
calculate block. In the manual mode, the measurement results of the first four calculate blocks
(Æ "CALCulate System", page 6.10) are shown in windows 1 to 4, while 8 calculate blocks are available
in the remote control mode.