Remote Control - Commands R&S NRP
1144.1400.12 6.122 E-3
Compatibility Information
This section contains information regarding the compatibility of the R&S NRP remote control commands
with those of the E4418B and E4419B power meters from Agilent.
The R&S NRP is largely downward-compatible to the above-mentioned devices but offers additional
capabilities.
Table 6-36: Compatibility information about remote control commands
E4418B/E4419B command Compatibility information
*DDT
Is not implemented in the R&S NRP.
CONFigure?
Returns the current configuration. The R&S NRP returns the last setting
made with
CONFigure, as specified by the SCPI standard.
CALCulate:GAIN
Is not implemented in the R&S NRP.
CALCulate:PHOLd
Is implemented in the R&S NRP and has the same effect as
CALC:EXTR:RES.
CALibration
:ECONtrol
:RCALibration
:RFACtor
Is not implemented in the R&S NRP.
DISPlay:ENABle[?]
Is implmented in the R&S NRP and has the same effect as
DISPlay:ILLUmination.
DISPlay:SCREen:FORMat
[:NUMeric[1]|2]
Is not implemented in the R&S NRP.
INITiate
:CONTinuous:SEQUence[1]|2
:IMMediate:SEQUence[1]|2
The :SEQUence node is not implemented in the R&S NRP.
MEMory:TABLe:GAIN
Agilent specifies the values without a unit; the unit PCT (percent) is
implicitly assumed. The units
DB (dB), DPCT (∆%) and O (unit 1) can
be specified in the R&S NRP. The default unit is
DB.
OUTPut:TRIGger:STATe
Is not implemented in the R&S NRP.
SENSe
:BANDwidth
:AVERage:COUNt:AUTO:SDETect
:DETector:FUNCtion
:MRATe
:SPEed
:V2P
Is not implemented in the R&S NRP.
SENSe[1|2]:CORRection
:CSET[1]|CSET2[:SELect]
and
SENSe[1|2]:CORRection
:CSET[1]|CSET2:STATe
The R&S NRP does not recognize any tables with frequency-
dependent calibration values since the sensors have to be calibrated
only once in production. For this reason, there is no compliance with
CSET1. Tables with frequency-dependent offset correction values
can be defined and addressed via :
CSET2 (Agilent and R&S NRP) or
.
:FDOTable (R&S NRP only).
SENSe[1|2]:CORRection:GAIN[1]?
Has the same effect as
SENSe[1..2]:CORRection:CFACtor?