Remote Control - Commands R&S NRP
1144.1400.12 6.44 E-3
OUTPut
The OUTPut system serves for generator configuration. The R&S NRP can be equipped with an
optional RF test generator which produces a reference power of 1 mW at 50 MHz. Moreover, the
analog outputs at the rear of the power meter can be configured.
Table 6-8: Commands of the OUTPut system
Command Parameter Unit Remark
OUTPut
:RECorder[1|2]
:FEED[?]
<string>
-
:LIMit
:LOWer[?]
<float_value>
-
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:UPPer[?] <float_value> -
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:STATe[?] ON|OFF -
:ROSCillator
[:STATe][?]
ON|OFF
-
Option R&S NRP-B1
:CALibration
:READ?
<NR1>
-
Option R&S NRP-B1
:WRIte <NR1>,<NR1> -
Option R&S NRP-B1
:TTL[1]
:ACTive[?]
LOW|HIGH
-
:FAIL[?] HIGH|LOW -
:FEED[?] <string> -
:HVOLtage <float_value> V
:LVOLtage <float_value> V
:STATe[?] ON|OFF -
OUTPut:RECorder[1..2]:FEED[?] <string>
Assigns a calculate block to one of the two analog outputs. On the basis of the set characteristic, the
measured value is converted to a DC voltage that can be tapped at the appropriate output.
Value range: "CALC1", "CALC2", "CALC3", "CALC4"
*RST value: output 1: "CALC1"
output 2: "CALC2"