Remote Control - Commands R&S NRP
1144.1400.12 6.30 E-3
DISPlay
The DISPlay command system is mainly used to configure the display of windows on the R&S NRP.
The WINDow commands are assigned to the calculate blocks 1 to 4. The values calculated from the
sensor values by the respective calculate block with the aid of the measurement function are displayed.
Table 6-5: Commands of the DISPlay system
Command Parameter Unit Remark
DISPlay
:ILLumination[?]
<boolean>
-
:ERRorlist[:STATe][?] <boolean> -
:MESSage
[:STATe][?]
-
-
:TEXT
:CLEar
-
-
No query
[:DATA] -
No query
:TYPE[?] QUERy|MESSage -
:PIXMap? - -
Query only
:UPDate[?] NORMal|SLOW|FREeze -
[:WINDow][1..4]
:AVALue[?] MINimum | MAXimum |
PTPeak | OFF
-
:FORMat[?] DIGital|ANALog|
GRAPhical
-
:METer | :ANALog
:AUTo[?] ONCE|OFF -
:LOWer[?] <float_value>
-
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:UPPer[?] <float_value> -
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:NAME[?] <string> -
:RESolution[?] 1|0.1|0.01|0.001 -
[:STATe][?] ON|OFF -
:TRACe
:LOWer[?]
<float_value>
DBM|
W|DBUV
:UPPer[?] <float_value> DBM|
W|DBUV
:TSLot[?] <int_value> -
:WINDow
:SELect[?] 1|2|3|4 -
:SIZE[?] NORMal|ZOOMed -
DISPlay:ILLumination[?] ON | OFF
3
This command switches the backlighting (ON) or (OFF). Backlighting is switched on again in case of a
reset.
*RST value: ON
3
DISP:ILL is identical to DISP:ENABle. See Annex "Compatibility Information".