R&S NRP Remote Control - Commands
1144.1400.12 6.31 E-3
DISPlay:ERRorlist[?] ON | OFF
Displays (ON) or removes (OFF) a windows which contains the contents of the SCPI error queue. The
queue may be deleted by pressing the ( DEL ) key.
*RST-Value: OFF
DISPlay:MESSage[:STATe][?] ON | OFF
Displays a window containing a user-defined message. The message text is defined via
DISP:MESS:TEXT. In addition, DISP:MESS:TYPE can select between two types of message windows.
Calling the command with the OFF parameter is ignored if DISP:MESS:TYPE QUERy is set, because in
this case a window must be closed by pressing the (¶MENU) key.
*RST value: OFF
DISPlay:MESSage:TEXT:CLEar
Deletes the stored text for user-defined messages.
DISPlay:MESSage:TEXT[:DATA] <string>
Defines the text that can be displayed in a message window. The text can contain ASCII characters
from code 32 (" ") to code 126 ("~"). The string "\n" is interpreted as a line break. Up to 4 lines
(DISP:MESS:TYPE MESS) or up to 2 lines (DISP:MESS:TYPE QUERy) can be shown. The length of a
line depends on the characters used, because a proportional font is used. Lines that are too long are
cut off, which is indicated by "..." at the end of the line.
*RST value: none
DISPlay:MESSage:TYPE QUERy | MESSage
Sets the mode for the expected user response, if the window with the user-defined message is
displayed:
QUERy
The message window must be closed by pressing the MENU) key. The execution of
remote control commands is blocked until this has been done.
MESSage
Remote control command processing is immediately continued. The message window
must be closed with DISP:MESS OFF if it is no longer needed.
*RST value: MESSage