R&S NRP Remote Control - Fundamentals
1144.1400.12 5.15 E-3
Annex: Automatic Settings for Manual Selection of a
Measurement Mode
When the measurement mode of a sensor is manually selected (SensorÆMode), automatic settings are
made for the trigger system and timing, which cannot be changed by manual control. This considerably
simplifies operation. These settings are not changed when the measurement mode is selected by
remote control (SENSe:FUNCtion) and have to be set to the desired value by appropriate SCPI
commands.
The following table lists the respective setting capabilities and the values that are automatically set in
manual control.
Table 5-2: Automatic settings for manual selection of a measurement mode
Mode SCPI command Value
Remark
TRIGger[1..4]:SOURce IMMediate
Cannot be modified in manual
mode.
TRIGger[1..4]:SLOPe -
Unchanged, is irrelevant.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay -
Unchanged, is irrelevant.
TRIGger[1..4]:HOLDoff 1 NS
Cannot be modified in manual
mode.
TRIGger[1..4]:LEVel -
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STARt -
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP -
Unchanged, is irrelevant.
ContAv
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.
TRIGger[1..4]:SOURce -
Unchanged, is irrelevant.
TRIGger[1..4]:SLOPe -
Unchanged, is irrelevant.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay -
Unchanged, is irrelevant.
Burst
TRIGger[1..4]:HOLDoff Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STOP Depends on sensor
User-selectable.
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.
TRIGger[1..4]:SOURce EXTernal | INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive | NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
in der HB nicht veränderbar
TRIGger[1..4]:DELay Depends on sensor
User-selectable.
TRIGger[1..4]:HOLDoff Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt -
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP -
Unchanged, is irrelevant.
Timegate
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.