Remote Control - Fundamentals R&S NRP
1144.1400.12 5.16 E-3
Mode SCPI command Value
Remark
TRIGger[1..4]:SOURce EXTernal | INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive | NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay Depends on sensor
User-selectable.
TRIGger[1..4]:HOLDoff Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt -
Unchanged, is irrelevant.
[SENSe[1..4]]:TIMing:EXCLude:STOP -
Unchanged, is irrelevant.
Timeslot
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.
TRIGger[1..4]:SOURce EXTernal | INTernal
User-selectable.
TRIGger[1..4]:SLOPe POSitive | NEGative
User-selectable.
TRIGger[1..4]:COUNt 1
Cannot be modified in manual
mode.
TRIGger[1..4]:DELay 0 S
Cannot be modified in manual
mode.
TRIGger[1..4]:HOLDoff Depends on sensor
User-selectable.
TRIGger[1..4]:LEVel Depends on sensor
User-selectable.
[SENSe[1..4]]:TIMing:EXCLude:STARt 0 S
Cannot be modified in manual
mode.
[SENSe[1..4]]:TIMing:EXCLude:STOP 0 S
Cannot be modified in manual
mode.
Scope
INITiate[1..4]:CONTinuous ON
Cannot be modified in manual
mode.