System settings (System menu) R&S NRP
1144.1400.12 4.58 E-2
Setting the analog outputs
There are two multifunction BNC connectors at the rear of the R&S NRP. The function of these connec-
tors is specified in the Analog Out dialog box.
System menu:
IO… – ( )
ä
Fig. 4-56 Analog Out dialog box
Analog … Pass/Fail – ( )
ä
Selects the function of port Out 1.
In the case of Analog, a voltage is output that is proportional to the
value currently displayed in the selected window (
Window 1...). In the
case of
Pass/Fail, a logic signal is output in the selected window (Window
1...
) to indicate that a limit value has been exceeded.
Remote Control, Analog
OUTPut:TTL[1]:STATe[?] OFF;OUTPut:RECorder[1..2]:STATe[?] ON
Pass/Fail
OUTPut:RECorder[1..2]:STATe[?] OFF;OUTPut:TTL[1]:STATe[?] ON
Window –
( )
1 2 3 4 ä
Specifies the window the output pertains to.
Remote Control
OUTPut:RECorder[1..2]:FEED[?] <string>
OUTPut:TTL[1]:FEED[?] <string>
0 V Equivalent –
( )
ä
Display value which corresponds to an output voltage of 0 V.
Remote Control
OUTPut:RECorder[1..2]:LIMit:LOWer[?] <float_value>
3 V Equivalent –
( )
ä
Display value which corresponds to an output voltage of 3 V.
Remote Control
OUTPut:RECorder[1..2]:LIMit:UPPer[?] <float_value>
Pass –
( )
ä
Logic level within the limits that have been set. The valid range is 0 V
to 3.3 V.
Remote Control
OUTPut:TTL[1]:HVOLtage[?] <float_value>
Fail –
( )
ä
Logic level when there is a limit violation. The valid range is 0 V to
3.3 V.
Remote Control
OUTPut:TTL[1]:LVOLtage[?] <float_value>
SystemIO…