R&S NRP Remote Control - Commands
1144.1400.12 6.5 E-3
SPD - Serial Poll Disable
Terminates the serial poll mode.
SPE - Serial Poll Enable
Activates the serial poll mode. When addressed as a talker, the device sends the status byte.
Common Commands to IEEE 488.2
The Common Commands to IEEE 488.2 are device commands that can be sent via the IEC/IEEE bus.
*CLS - CLear Status
Clears the status of the R&S NRP by resetting the following registers:
Status byte register
Standard event status register
Error / event queue
All SCPI status registers
Note: The enable and transition registers (NTRansition and PTRansition) are not
changed!
*ESE 0..255 - Standard Event Status Enable
Sets the ENABle register of the Standard Event Status Register (ESR) to the specified value(Æ SCPI
command STATus, page 6.68).
*ESE? - Standard Event Status Enable Query
Returns the content of the ENABle register of the Standard Event Status RegisterÆ SCPI command
STATus , page 6.68).
*ESR? - Standard Event Status Register Query
Returns the content of the EVENt register of the Standard Event Status RegisterÆ SCPI command
STATus , page 6.68).
*IDN? - IDeNtification Query
Returns a string with information on the R&S NRP identity (device identification code). The string has
the following format <string1>,<string2>,<string3>,<string4>.