Graphically representing power versus time (Scope mode) R&S NRP
1144.1400.12 2.18 E-2
Graphically representing power versus time (Scope mode)
In the Scope mode, power versus time can be represented graphically as with an oscilloscope. If you
wish to carry out time-dependent measurements (in the Burst or Timeslot mode) on modulated signals,
displaying the signal in the Scope mode is always initially recommended since stable triggering by the
R&S NRP is most clearly recognized in this mode.
1 Setting the Scope mode.
¾ Connect a power sensor of the R&S NRP-Z1x or
R&S NRP-Z2x family to connector A of the R&S NRP
and apply an amplitude-modulated or pulsed signal
having a modulation or pulse frequency of approx.
1 kHz. The maximum level should be between
–10 dBm and +10 dBm.
({PRE}SET)
* *
¾ Press the ({PRE}SET) key twice.
A measurement window is now available.
( )
**
¾ Change to the Sensor menu and select Mode.
Mode … ( )
*
… T'slot T'gate Scope ( )
**
¾ Switch to the Scope mode.
The associated parameters will appear in the lower part
of the dialog box. You can leave the parameters un-
changed for this exercise.
For a detailed description, see Chapter 4, Displaying data
in the Scope mode.
(¶ MENU)
*
¾ Close the Mode dialog box.
( )
**
¾ Change to the Windows menu and press Expand.