Remote Control - Commands R&S NRP
1144.1400.12 6.66 E-3
SERVice
The commands of the SERVice system are used for maintenance or test purposes. They were
implemented primarily to support device development. The numeric suffix refers to the sensors.
Note: Before commands of the SERVice system can be sent to a sensor, they
must be enabled via the SERVice:UNLock 1234 command.
Table 6-12: Commands of the SERVice command system
Command Parameter Unit Remark
SERVice[1..4]
:CALibration
:TEST? <NR1> -
Query only
:TEMP ONCE -
No query
:TEMP:DATA? <NR1 > Kelvin
Query only
:RCOunt[?] <NR1> -
:RESult[?] <NR1> -
:UNLock[?] 1234 -
SERVice[1..4]:CALibration:TEST? <NR1>
Tests the consistency of the calibration data set according to three criteria:
If the following bit is set in <NR1>, the following action is executed.
0 The serial number of the calibration data set is tested.
1 The checksum of the calibration data set is tested.
2 The header of the calibration data set is tested.
If an inconsistency of the data set is detected, the query returns a 1, otherwise a 0.
Value range: 0 to 255
*RST value: none
SERVice[1..4]:CALibration:TEMP ONCE
Initiates a temperature measurement of the sensor in question. The temperature can be queried with
SERV:CAL:TEMP:DATA?.
SERVice[1..4]:CALibration:TEMP:DATA? <NR1>
Returns the temperature of the sensor in Kelvin if a temperature measurement was previously initiated
by means of SERV:CAL:TEMP ONCE.