R&S NRP Remote Control - Commands
1144.1400.12 6.67 E-3
SERVice[1..4]:RCOunt[?] <NR1>
SERVice[1..4]:RESult[?] <NR1>
These commands allow simulated measured values to be specified for test purposes. SERV:RCO
specifies the number of SERV:RES pairs to follow. The first SERV:RES command specifies the
measured value in watts, the second SERV:RES command determines how often the measured value
should be obtained. The measurement sequences defined in this way are repeated until the sensor is
disconnected from the R&S NRP and reconnected.
Example:
The command sequence
SERV2:RCO 2
SERV2:RES 3
SERV2:RES 2
SERV2:RES 4
SERV2:RES 1
makes sensor 2 return the measured values 3 W, 3 W, 4 W (, 3 W, 3 W, 4W, ....) from now on.
*RST value: none. The simulation mode is not exited in the case of *RST.
SERVice[1..4]:UNLock[?] 1234
Before other SERVice system commands are accepted, they must be enabled with SERV:UNL 1234.