1144.1400.12 6.108 E-3
UNIT
Table 6-35: Commands of the UNIT system
Command Parameter Unit Remark
UNIT[1..8]
:POWer[?]
DBM | W
-
:RATio[?] DB | PCT
-
UNIT[1..8]:POWer[?] DBM | W | DBUV
Selects the output unit for the measured power values. This setting also determines the unit for the
parameters of the following commands:
CALC:REL
CALC:LIM:UPP
CALC:LIM:LOW
DISP:MET:UPP
DISP:MET:LOW
OUTP:REC:LIM:LOW
OUTP:REC:LIM:UPP
*RST value: DBM
UNIT[1..8]:POWer:RATio[?] DB | DPCT | O
Selects the output unit for the measured power ratio values. This setting also determines the unit for the
parameters of the following commands:
CALC:REL
CALC:LIM:UPP
CALC:LIM:LOW
DISP:MET:UPP
DISP:MET:LOW
OUTP:REC:LIM:LOW
OUTP:REC:LIM:UPP
*RST value: DB
Error messages: If the compute function of the associated calculate block is :SWR, :RLOSs or
:REFLection, then the unit cannot be changed and error message 26,"State not supported", is output.