Configuring measurements (Measurement menu) R&S NRP
1144.1400.12 4.40 E-2
Limit monitoring
An upper and a lower limit for each window can be set in the Limits… dialog box. If either limit is vio-
lated, a warning is issued. The warning may be one of the following:
The LIMIT FAIL message is displayed on the screen.
A warning tone is output.
A signal is issued on analog output OUT1 ( Setting the analog outputs on page 4.58).
Measurement menu:
Limits… – ( )
ä
Fig. 4-34 Limits dialog box
Upper [ß ]( )
ä
Turns monitoring of the upper limit on/off.
Remote Control
CALCulate[1..8]:LIMit:UPPer:STATe[?] ON | OFF
[ .... ] Upper ( )
ä
Upper limit.
Remote Control
CALCulate[1..8]:LIMit:UPPer[:DATA][?] <float_value>
Lower
[ß ]( )
ä
Turns monitoring of the lower limit on/off.
Remote Control
CALCulate[1..8]:LIMit:LOWer:STATe[?] ON | OFF
[ .... ] Lower ( )
ä
Lower limit
Remote Control
CALCulate[1..8]:LIMit:LOWer[:DATA][?] <float_value>
Warning Beep
( )
ää
For activating the acoustic alarm.
Remote Control
CALCulate[1..8]:LIMit:BEEP[?] ON | OFF
MeasurementLimits