EasyManua.ls Logo

Rohde & Schwarz R&S NRP - Calculate (Configuration of Calculate Blocks)

Rohde & Schwarz R&S NRP
350 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S NRP Remote Control - Commands
1144.1400.12 6.21 E-3
CALCulate (Configuration of Calculate Blocks)
The calculate blocks calculating the measurement results from the sensor data are configured with the
CALCulate command. Eight blocks are available, each with two input channels (primary and
secondary channel). The available sensors (1 to 4, depending on device configuration and connected
sensors) can be assigned to the channels. With the aid of a selectable function (CALC:MATH), the
channels can be processed together.
Table 6-2: Commands of the CALCulate system
Command Parameter Unit Remark
CALCulate[1..8]
:DATA?
-
Query only.
The unit is the current unit of the
measurement result.
:MINimum:DATA? - -
Query only.
The unit is the current unit of the
measurement result.
:MAXimum:DATA? - -
Query only.
The unit is the current unit of the
measurement result.
:PTPeak:DATA? - -
Query only.
The unit is the current unit of the
measurement result.
:EXTRemes:RESet - -
No query
:FEED[1..2][?] <string> -
:LIMit
:BEEP[?]
ON|OFF
-
No query
:CLEar[:IMMediate] - -
No query
:AUTO[?] - -
:FAIL? - -
Query only
:FCOunt? - -
Query only
:UPPer
[:DATA][?] <float_value> -
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:STATe[?] ON|OFF -
:LOWer
[:DATA][?] <float_value> -
The value range of the parameter
depends on the output unit of the
measured value.
:POWer[?] <float_value> DBM|
W|DBUV
:RATio[?] <float_value> DB|
DPCT|O
:STATe[?] ON|OFF -
:MATH
[:EXPRession][?] <string> -
:CATalog? - -
Query only
:RELative
[:MAGNitude][?] <float_value> -
The value range of the parameter
depends on the output unit of the
measured value.
:AUTO[?] OFF|ONCE -
:POWer[:MAGNitude][?] <float_value> DBM|
W|DBUV
:RATio[:MAGNitude][?] <float_value> DB|
DPCT|O
:STATe[?] ON|OFF -

Table of Contents

Related product manuals