Remote Control - Commands R&S NRP
1144.1400.12 6.20 E-3
<dtolerance>
Value range: depending on sensor.
Unit: S
Default value: depending on sensor.
The length of a time interval in which the power level may drop below the
trigger level without the end of a power pulse being detected.
Æ[SENSe[1..4]][:POWer]:BURSt:DTOLerance[?]
<float_value>
<capture_time>
Value range: depending on sensor.
Unit: S
Default value: depending on sensor.
Period within which measured data are captured in the Scope mode.
Æ [SENSe[1..4]]:TRACe:TIME[?] <float_value>
(<buffered_size>)
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
Number of requested measured values.
Æ [SENSe[1..4]][:POWer][:AVG]:BUFFer:SIZE[?]
<int_value>
(<scope_size>)
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
Number of measured values in one data set.
Æ [SENSe[1..4]]:TRACe:POINts <int_value>
Error messages
24
"Sensor mode not supported": A sensor does not support a measurement mode.
-221
"Settings conflict": This error is generated when a FETCH or READ command is sent and the
current device settings do not correspond to the transmitted parameters. The parameters
transmitted in these two commands are not used for device configuration but rather to check the
settings.
-108
"Parameter not allowed": The parameter list contains unexpected parameter types or the
parameters are not sent in the correct order.
-127
"Invalid numeric data": An invalid value was specified for <resolution>.
-171
"Invalid expression": Error in the <source_list> parameter. Two sensors were specified
instead of one, or the specified sensor is not correct.