Remote Control - Commands  R&S NRP 
1144.1400.12 6.20  E-3 
 
<dtolerance> 
Value range: depending on sensor. 
Unit: S 
Default value: depending on sensor. 
The length of a time interval in which the power level may drop below the 
trigger level without the end of a power pulse being detected. 
Æ[SENSe[1..4]][:POWer]:BURSt:DTOLerance[?] 
<float_value> 
 
<capture_time> 
Value range: depending on sensor. 
Unit: S 
Default value: depending on sensor. 
Period within which measured data are captured in the Scope mode. 
Æ [SENSe[1..4]]:TRACe:TIME[?]  <float_value> 
 
(<buffered_size>) 
Value range: depending on sensor. 
Default value: depending on sensor. 
Unit: - 
Number of requested measured values.  
Æ [SENSe[1..4]][:POWer][:AVG]:BUFFer:SIZE[?] 
<int_value> 
 
(<scope_size>) 
Value range: depending on sensor. 
Default value: depending on sensor. 
Unit: - 
Number of measured values in one data set. 
Æ [SENSe[1..4]]:TRACe:POINts  <int_value> 
 
 
 
 
Error messages 
24 
"Sensor mode not supported": A sensor does not support a measurement mode. 
-221 
"Settings conflict": This error is generated when a FETCH or READ command is sent and the 
current device settings do not correspond to the transmitted parameters. The parameters 
transmitted in these two commands are not used for device configuration but rather to check the 
settings. 
-108 
"Parameter not allowed": The parameter list contains unexpected parameter types or the 
parameters are not sent in the correct order. 
-127 
"Invalid numeric data": An invalid value was specified for <resolution>. 
-171 
"Invalid expression": Error in the <source_list>  parameter. Two sensors were specified 
instead of one, or the specified sensor is not correct.