R&S NRP Remote Control - Commands
1144.1400.12 6.19 E-3
Æ [SENSe[1..4]]:AVERage:COUNt:AUTO:RESolution[?] 1 | 2
| 3 | 4,
Æ [SENSe[1..4]]:TRACe:AVERage:COUNt:AUTO:RESolution[?]
1|2|3|4 and
Æ DISPlay[:WINDow][1..4]:RESolution[?] 1 | 0.1 | 0.01
| 0.001
Example:
If
<resolution> is set to 3 or 0.01, three significant points are assigned
to the value when it is displayed on a linear scale. When a log scale is
used, 2 places are displayed after the decimal point.
<source_list>
Value range: (@n) | (@n),(@m)
Default value: (@1) or (@1),(@2)
n and m may assume the values 1, 2, 3 or 4.
With this list, sensors are assigned to the primary and possibly to the
secondary channel of a calculate block. The number for each sensor is
preceded by the character @ and the entire expression is enclosed in
parentheses.
Example:
(@1)is used for sensor A.
Example:
The values (@3),(@2) are set for <source_list>. Sensor C is
assigned to the primary channel and sensor B to the secondary channel.
The calculate blocks process the primary channel together with the
secondary depending on the desired function. If
:RATio is selected as the
calculation function, the result from the calculate block is the ratio of the
values measured by sensor C and sensor B.
<tslot_width>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: S
Width of a timeslot in the Timeslot mode.
Æ [SENSe[1..4]][:POWer]:TSLot:[AVG]:WIDTh[?]
<float_value>
<no_slots>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
Number of timeslots to be measured in the Timeslot mode.
<start_exclude>
<end_exclude>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: S
Amount of time at the beginning and the end of a timeslotor integration
period that should not be taken into account.
Æ [SENSe[1..4]]:TIMing:EXCLude:STARt[?] <float_value>
and
Æ [SENSe[1..4]]:TIMing:EXCLude:STOP[?] <float_value>