R&S NRP Remote Control - Commands
1144.1400.12 6.107 E-3
TRIGger[1..4|:ALL]:SOURce[?]
BUS | EXTernal | HOLD | IMMediate | INTernal
Sets the trigger signal source for the WAIT_FOR_TRG state.
BUS
The trigger event is initiated by TRIG:IMM or *TRG. In this case, the other trigger setting
are meaningless.
EXTernal
Triggering is performed with an external signal applied to the trigger connector. The
TRIG:SLOP command determines whether the rising or the falling edge of the signal is
to be used for triggering. Waiting for a trigger event can be skipped by TRIG:IMM.
IMMediate
The sensor does not remain in the WAIT_FOR_TRG state but immediately changes to
the MEASURING state.
HOLD
A measurement can only be triggered when the command TRIG:IMM is executed.
INTernal
The sensor determines the trigger time by means of the signal to be measured. When
this signal exceeds (TRIG:SLOP POS) or fall short of (TRIG:SLOP NEG) the power set
by TRIG:LEV, the measurement is started after the time set by TRIG:DEL. Similar to
TRIG:SOUR EXT, waiting for a trigger event can also be skipped by TRIG:IMM.
*RST value: depending on sensor