Remote Control - Commands R&S NRP
1144.1400.12 6.106 E-3
TRIGger[1..4|:ALL][:IMMediate]
Performs triggering and ensures that the sensor directly changes from the WAIT_FOR_TRG state to
the MEASURING state irrespective of the selected trigger source (
TRIG:SOUR). A trigger delay set with
TRIG:DEL is ignored but not the automatic delay determined when TRIG:DEL:AUTO:ON is set.
When the trigger source is HOLD, a measurement can only be started with TRIG.
Error messages:
-211
"Trigger ignored": The sensor is not in the WAIT_FOR_TRG state.
TRIGger[1..4|:ALL]:LEVel[?] <float_value>
Determines the power a trigger signal must exceed (TRIG:SLOP POS) or fall short of (TRIG:SLOP
NEG) before a trigger event is detected. This setting is only used for the trigger signal source
(TRIG:SOUR) INTernal.
Unit: DBM | W
Default unit: DBM
Value range: depending on sensor
*RST value: depending on sensor
TRIGger[1..4|:ALL]:SLOPe[?] POSitive | NEGative
This command determines whether the rising (POSitive) or the falling (NEGative) edge of the signal
is used for triggering.
*RST value: depending on sensor