R&S NRP Remote Control - Commands
1144.1400.12 6.105 E-3
TRIGger[1..4|:ALL]:COUNt[?] <int_value>
Sets the number of measurement cycles to be performed when the measurement is started with INIT.
Unit: 1
Value range: depending on sensor
*RST value: depending on sensor
TRIGger[1..4|:ALL]:DELay[?] <float_value>
Defines the delay between the trigger event and the beginning of the actual measurement (integration).
Unit: S
Value range: depending on sensor
*RST value: depending on sensor
TRIGger[1..4|:ALL]:DELay:AUTO[?] ON | OFF
TRIG:DEL:AUTO ON ensures by means of an automatically determined delay that a measurement is
started only after the sensor has settled. This is important when thermal sensors are used. The
automatically determined delay is ignored when a longer period was set with TRIG[1..4]:DEL.
*RST value: depending on sensor
TRIGger[1..4|:ALL]:HOLDoff[?] <float_value>
Defines a period after a trigger event within which all further trigger events are ignored.
Unit: S
Value range: depending on sensor
*RST value: depending on sensor
TRIGger[1..4|:ALL]:HYSTeresis[?] <float_value>
This command is used to specify how far the signal level has to drop below the trigger level before a
new signal edge can be detected as a trigger event. Thus, this command can be used to eliminate the
effects of noise in the signal on the edge detector of the trigger system.
Unit: DB | PCT
Default unit: DB
Value range: depending on sensor
*RST value: depending on sensor