R&S NRP Remote Control - Commands
1144.1400.12 6.13 E-3
Syntactical Structure of High-Level Measurement Commands
The high-level measurement commands can be divided as follows:
<head><function>[<expression>][?] <parameter_list>
<head>
CONFigure[1..8], READ[1..8], MEASure[1..8], FETCh[1..8]
<function>
Selection of one of the measurement modes ContAv, ContAv with data
buffering, Burst, Timeslot or Scope. (ÆSENSe[1..4]:FUNCtion command
on page 6.56).
Keywords:
• [:SCALar][:POWer][:AVG]
• [:SCALar][:POWer]:TSLot
• [:SCALar][:POWer]:BURSt
• :ARRay[:POWer][:AVG]
• :XTIMe[:POWer]
<expression>
Specification of a calculation function for the addressed CALCulate block (Æ
CALCulate[1..8]:MATH:EXPRession command on page 6.26).
If <expression> is not specified, the measured value of the primary channel
is output.
Keywords:
• :RELative
• :DIFFerence
• :DIFFerence:RELative
• :SUM
• :SUM:RELative
• :RATio
• :RATio:RELative
• :RLOSs
• :SWR
• :REFLection
[?]
Queries must be terminated with an "?".
<parameter_list>
A list with parameters whose meaning is determined by the selected
<function>.
Since all four high-level measurement commands use the same parameter lists, they are described
together in this section.