R&S NRP Remote Control - Commands
1144.1400.12 6.17 E-3
:SWR
PCSC
PCSC
/1
/1
+
If the forward power of a wave is measured in the PC and the
reflected power in the SC, the measurement command yields the
standing wave ratio. The output unit is set to "One" (UNIT:RAT
O).
:REFLection
PCSC /
If the forward power of a wave is measured in the PC and the
reflected/transmitted power in the SC, the calculate block yields
the reflection coefficient/transmission factor of the DUT. The
output unit is set to "One" (UNIT:RAT O).
:RLOSs
PCSC /log20
10
If the forward power of a wave is measured in the PC and the
reflected or transmitted power in the SC, the calculate block
yields the return loss/transmission loss of the DUT.
This function principally yields the same result as the :RATio
function. The only difference is the output unit which is
automatically set to dB by means of this command (UNIT:RAT
DB).