R&S NRP Remote Control - Commands
1144.1400.12 6.23 E-3
CALCulate[1..8]:LIMit:BEEP ON | OFF
Switches the acoustic warning for limit violations on or off.
*RST value: OFF
CALCulate[1..8]:LIMit:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
CALCulate[1..8]:LIMit:CLEar:AUTO ON | OFF | ONCE
Automatically resets the limit monitoring state and the internal counter for limit violations if any of the
following events occurs:
INITiate[:IMMediate] is executed
INITiate:CONTinuous ON is executed
A MEASure? or MEASure:...? query is executed.
A READ? or READ:...? query is executed.
If the automatic function is set to ONCE, the counter is reset only when the next of these events occurs.
*RST value: OFF
CALCulate[1..8]:LIMit:FAIL?
Queries whether upper or lower limits have been exceeded. The status is reset by the following events
(0 = no limit violation):
The device is switched on
A reset is performed (*RST)
The CALCulate[1..8]:LIMit:CLEar command is sent.