R&S NRP Remote Control - Commands
1144.1400.12 6.25 E-3
CALCulate[1..8]:LIMit:LOWer:STATe[?] ON | OFF
This command switches the monitoring function for the lower limit on or off.
*RST value: OFF
CALCulate[1..8]:LIMit:UPPer[:DATA][?] <float_value>
This command sets an upper limit for the measured values. The R&S NRP can respond when this limit
is exceeded (see other
CALCulate[1..8]:LIMit commands).
The value range depends on the current output unit of the measured value of the calculate block.
Unit of
measured
value
W DBM DBUV DPCT DB O
Value range
-9.99999E5
..
9.99999E5
-120
..
90
-13.0103
..
196.897
-100
..
999.99
-200
..
200
-1E20
..
1E20
Unit: Æ Table 6-3 on page 6.27.
*RST value: 0 DBM or 0 DB
CALCulate[1..8]:LIMit:UPPer[:DATA]:POWer[?] <float_value>
This command sets an upper limit for the measured power (unit W, DBM or DBUV). The R&S NRP can
respond when this limit is exceeded (see other CALCulate[1..8]:LIMit commands).
Unit:
DBM | W | DBUV
Default unit:
DBM
Value range: -120 DBM to 90 DBM
*RST value: 0 DBM
CALCulate[1..8]:LIMit:UPPer[:DATA]:RATio[?] <float_value>
This command sets an upper limit for the measured power ratios (unit DB, DPCT or O). The R&S NRP
can respond when this limit is exceeded (see other CALCulate[1..8]:LIMit commands).
Unit:
DB | DPCT | O
Default unit:
DB
Value range: -200 DB to 200 DB
*RST value: 0 DB
CALCulate[1..8]:LIMit:UPPer:STATe[?] ON | OFF
This command switches the monitoring function for the upper limit on or off.
*RST value: OFF