R&S NRP Remote Control - Commands
1144.1400.12 6.57 E-3
[SENSe[1..4]][:POWer][:AVG]:APERture[?] <float_value>
This command determines the integration time for a single measurement in the ContAv mode. To
increase the measurement accuracy, this integration is followed by a second averaging procedure in a
window with a selectable number of values. The filter window is configured with the
[SENSe[1..4]]:AVERage commands.
Unit: S
Value range: depending on sensor
*RST value: depending on sensor
[SENSe[1..4]][:POWer][:AVG]:BUFFer:SIZE[?] <int_value>
Sets the number of desired values for the buffered ContAv mode.
Unit: -
Value range: 1..400000
*RST value: depending on sensor
[SENSe[1..4]][:POWer][:AVG]:BUFFer:STATe[?] ON | OFF
Switches on the buffered ContAv mode, after which data blocks rather than single measured values are
then returned. In this mode a higher data rate is achieved than in the non-buffered ContAv mode.
The number of desired measured values is set with the SENSe[1..4]:POWer:ARR:SIZE command.
*RST value: depending on sensor
[SENSe[1..4]][:POWer][:AVG]:RANGe[?] 0 | 1 | 2
Selects a measurement range in which the corresponding sensor is to perform a measurement. This
setting will only become effective if SENS[1..4]:RANG:AUTO ON is set.
*RST value: depending on sensor
[SENSe[1..4]][:POWer][:AVG]:RANGe:AUTO[?] ON | OFF
Sets the automatic selection of a measurement range to ON or OFF.
*RST value: depending on sensor
[SENSe[1..4]][:POWer][:AVG]:RANGe:AUTO:CLEVel[?] <float_value>
Sets the cross-over level. Shifts the transition ranges between the measurement ranges. This may
improve the measurement accuracy for special signals, i.e. signals with a high crest factor.
Unit: DB | PCT
Default unit: DB
Value range: depending on sensor
*RST value: depending on sensor