SYSTem:PRESet
Resets the R&S NRP to default (*RST) values. No difference is made between reset and preset values
(exceptions: INIT:CONT, SENS:AVER:TCON und SENS:TRAC:AVER:TCON). For preset values see
Table 6-33 .
Table 6-33: Preset and *RST value
Command Preset and *RST value Remark
CALC[1..8]:FEED[1..2] "POW:AVER"
The calculate block processes
average power values ("Average").
CALC[1..8]:LIM:BEEP OFF
No audible signal in case of limit
violations.
CALC[1..8]:LIM:CLE:AUTO OFF
No automatic reset of the limit
monitoring function when a
measurement is started.
CALC[1..8]:LIM:FAIL? 0
Limit violation.
CALC[1..8]:LIM:FCO? 0
Number of limit violations is reset.
CALC[1..8]:LIM:UPP:DATA
0 W or 0 DB
Upper limit value for measurements.
CALC[1..8]:LIM:UPP:STAT OFF
No monitoring function for upper limit
value.
CALC[1..8]:LIM:LOW:DATA
0 W or 0 DB
Lower limit value for measurements.
CALC[1..8]:LIM:LOW:STAT OFF
No monitoring function for lower limit
value.
CALC[1..8]:MATH
Æ For command
description see 6.26
The calculate block returns the
measured value of the sensor
assigned to the primary channel.
CALC[1..8]:REL
0 DBM or 0 DB
-
CALC[1..8]:REL:POW 0 DBM
-
CALC[1..8]:REL:RAT 0 DB
-
CALC[1..8]:REL:STAT OFF
Measurements are not converted into
a reference value.
DISP:ILL -
Display lighting remains unchanged
and can be activated by pressing
( Æ ).
DISP[1..4]:AVAL OFF
Dont show additional information on
measured values.
DISP:MESS OFF
Do not display messages.
DISP:MESS:TEXT -
No change.
DISP:MESS:TYPE MESS
Messages cannot be confirmed
manually.
DISP[1..4]:FORM DIG
Digital result display.
DISP[1..4]:MET:LOW -90 DBM
-120 DB
Lower limit value for analog result
display.
DISP[1..4]:MET:UPP 70 DBM
60 DB
Upper limit value for analog result
display.
DISP[1..4]:NAME ""
Text to be displayed in result window.
DISP[1..4]:RES 0.01
Measurement accuracy and display
accuracy.
DISP:SEL 1
Selection of first result window.
DISP:SIZE NORM
Automatic window sizes.
DISP[1..4] ON | OFF
A window will automatically be
opened for each sensor.
DISP[1..4]:TRAC:LOW -120 DBM
Lower limit of power axis in Scope
mode