R&S NRP Window handling
1144.1400.12 2.15 E-2
3 Auxiliary values (Maximum, Minimum, Max, Min, …).
( )
* *
¾ Select the Measurement menu.
Function… ( )
*
¾ Select Function…in this menu.
The measurement function, unit, sensors used and the
auxiliary value can be selected in the Function & Unit
dialog box.
Auxiliary Value ( )
* *
¾ Open the Auxiliary Value drop-down list.
Auxiliary Value ( )
* *
¾ Select Max.
(¶ MENU)
*
¾ Close the dialog box.
¾ Press the Max Hold Reset softkey and close the menu
with (¶ MENU).
The updated maximum is now displayed on the right next
to the measured value.
If you reduce the signal power, the maximum should
remain the same. However, if the signal power is in-
creased, the maximum changes accordingly.