132 Operating Instructions Combics CAIXS2
Menu Structure
setup /com-1 dat.prot Data protocols
Config. SBI*
BAUD Baud rate 5.1
150 150 5.1.1
300 300 5.1.2
600 600 5.1.3
1200 1200 5.1.4
2400 2400 5.1.5
4800 4800 5.1.6
9600 9600 5.1.7*
19200 19200 5.1.8
PARITY Parity 5.2
SPACE Space
Only if 7 data bits is selected 5.2.2
ODD Odd 5.2.3*
EVEN Even 5.2.4
NONE None 5.2.5
STOPBIT Number of stop bits 5.3
1 STOP 1 stop bit 5.3.1*
2 STOP 2 stop bits 5.3.2
HANDSHK Handshake mode 5.4
SOFTW. Software handshake 5.4.1
HARDW. Hardware handshake,
1 character after CTS 5.4.3*
DATABIT Number of data bits 5.6
7 bits* 5.6.1
8 bits 5.6.2
MAN./AUT. Data output (manual/automatic) 6.1
IND.W/O Manual, without stability 6.1.1
IND.AFTR Manual, at stability 6.1.2*
AUT. W/O Automatic, without stability 6.1.4
AUT.WITH Automatic, with stability 6.1.5
PROT.PRN Protocol printout for
computer (PC) 6.1.7
AUT.CYCL Time-dependent automatic data output 6.3
EACHVAL 1 display update 6.3.1*
AFTR.2 2 display updates 6.3.2
AFTR.10 10 display updates 6.3.4
AFTR.100 100 display updates 6.3.7
LINE Data output: Line format 7.2
16. CHAR For raw data: 16 characters 7.2.1
22. CHAR For other applications:
22 characters 7.2.2*
SIGN Data output: Sign format 7.3
+ DEACT. Plus sign deactivated 7.3.1
+ ACT. Plus sign activated 7.3.2*
SETTING Factory settings for COM1: SB 9.1
YES Yes 9.1.1
NO No* 9.1.2
xBPI-232
SMA
BAUD Baud rate 5.1
150 150 5.1.1
300 300 5.1.2
600 600 5.1.3
1200 1200 5.1.4
2400 2400 5.1.5
4800 4800 5.1.6
9600 9600 5.1.7*
19200 19200 5.1.8
Numeric menu 5.2 to 5.6 similar to SBI
* = Factory setting