Operating Instructions Combics CAIXS2 147
Verification of Intrinsic Safety
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
6
8
66015-751-60
RS232 Data Interface
Non-hazardous area
Hazardous area
RS232 data interface
(pins A/J/K/N and M)
1)
Ui 12.6 V Uo 12.6 V
Ii 85 mA Io 28 mA
Pi 270 mW Po 88 mW
Ci 3 nF Co 1.15 µF
Li 0 Lo 50 mH
Note:
1: Per circuit
2: Combined circuits
3: Not all models can be used in zone 21,22
4: BAS01ATEX7005; II (1) GD [EEx ia] IIC
5: Both channels connected on Z966
*: Versus GND; **: Between the lines; ***: Resistively limited
YDI05-Z.. interface converter
II (2) GD [EEx ib] IIC or
Z966 Zener barrier
4)
in YDI02-Z..: II (2) G [EEx ib] IIC or
YCO01-Y interface converter
II (2) GD [EEx ib] IIC or
II 3 (2)GD EEx nR[ib]IIC T4
6-wire standard cable, up to 0.5mm² stranded wire, with
max. 250nF/km, 750µH/km and minimum 34ohm/km has
22µH/ohm. However, the length of the cable (flexibly
installed) is limited to under 25m by the RS-232
specifications. 6-wire cable, type PR6136 (Lmax =
1.1mH/km; Cmax = 220nF/km; Rmin = 26 ohms/km:
43µH/ohm): under 25m (see above).
Alternative
connections
YDI05-Z..
2)
Uo 12.4 V* Ui 12.6 V*
24.8 V** 25.2 V**
Io 260 mA*** Ii any
Po 800 mW Pi any
Co 1.24 µF* Ci 0
112 nF**
Lo 400 µH* Li 0
400 µH**
Lo/Ro 44 µH/
* / 22 µH/ **
Z966
1)
pins A/J/K/N and M
Uo 12 V*
24 V**
Io 82 mA /164 mA
8)
Po 0.24 W / 0.48 W
8)
Co 1.41 µF*
125 nF**
Lo 5.52 mH
1.38 mH
8)
Lo/Ro 147 µH/
57 µH/
8)
YCO01-Y
2)
Uo 11.8 V* Ui 12.6 V*
23.6V** 25.2 V**
Io 123 mA*** Ii 131 mA
Po 361 mW Pi any
Co 1.5 µF* Ci 0.5 nF
129 nF**
Lo 2 mH* Li 0.8 µH
2 mH**
Lo/Ro 98 µH/
* / 98µH/ **
SIWS… / ISXS.... / CAIXS2 RS232 data interface
2)
COM1
Ui 12.6 V* / 25,2V** Uo 10 V* / 20V**
Ii 328 mA*** Io 101 mA***
Pi any Po 253 mW
Ci 2.2 nF* / 0.5nF** Co 3 µF* / 217nF**
Li 0 mH Lo 3 mH
Lo/Ro 140 µH/
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
FC/FCA/IS.......-.X..
3)
scale / weighing platform
SIWXS…-3-..-... /
ISX...-..-... / CAIXS2
3)
scale / weighing platform
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
5
8
66015-751-60
Digital I/Os
Non-hazardous area
Hazardous area
Safety barrier
1)
Uo 8.6 V
Io 50 mA
Po 107.5 mW
Co 6.2 µF
Lo 15 mH
Foot switch
3)
9001/01-086-050-101
Safety barrier
(supplied by Stahl)
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
Junction box
3)
Digital I/Os (COM1)
2)
Ui 8.6 V Uo 6,0 V
Ii -- Io 45 mA
Pi -- Po 67 mW
Ci 0 nF Co 40 µF
Li 0 mH Lo 20 mH
Lo/Ro 530 µH/
Switch
3)
Alternative
connection
Cable with up to 10 wires and with max.
250nF/km and 1.1mH/km: maximum cable
length 1000m (flexibly installed).
Note:
1: Per circuit
2: Combined circuits
3: Suitable equipment may be optionally used; only passive wiring/switch
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db

Table of Contents