EasyManua.ls Logo

Sartorius Combics 2 CAIXS2 - Verification of Intrinsic Safety; RS485 Data Interface

Sartorius Combics 2 CAIXS2
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
148 Operating Instructions Combics CAIXS2
Verification of Intrinsic Safety
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
8
8
66015-751-60
RS422 Data Interface
Non-hazardous
area
Hazardous area
YCO01-Y
Interface converter
Note:
1: Per circuit
2: Combined circuits
*: Versus GND; **: Between the lines; ***: Resistively limited
Cable:
For a standard cable (e.g., type LiYCY with max. 250nF/km
(wire versus wire or wire versus shield), max. 750µH/km and
at least 34 ohms/km (e.g. 0.5mm² stranded wire), this yields
Lo/Ro = 22 µH/ohm
Permitted length of the flexibly installed
cable: 330m
YDI05-Z
(Option A25)
Interface converter
II (2) GD [EEx ib] IIC
Alternative
connection
RS422 data interface
2)
(COM1)
Uo 5,2 V Ui 8.6 V
Io 290 mA Ii 210 mA
Po 496 mW Pi 0.5 W
Co 60 µF Ci 0.5 µF
Lo 300 µH Li 0 mH
Lo/Ro 50 µH/
RS422 data interface of YCO01-Y
2)
Uo 7.2 V* Ui 12.6 V*
8.0 V** 25.2 V**
Io 207 mA*** Ii 1.5 A
Po 330 mW Pi 2.5 W
Co 13.5 µF* Ci 1 nF
8.4 µF**
Lo 0.7 mH* Li 1.6 µH
0.7 mH**
Lo/Ro 36 µH/
*/ 36µH/ *
RS422 data interface of YDI05-Z
2)
Uo 6.0 V *
)
Ui 12.6 V *
)
6.8 V **
)
25.2 V **
)
Io 172 mA (linear) Ii 0.2 A
Po 0.5 W Pi 2.3 W
Co 12 µF *
)
Ci 28 µF *
)
Co 17.7 µF
**)
Ci 200 nF **
)
Lo 0.7 mH Li 2 µH
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC
Verification of
Intrinsic Safety
CAIXS2 /
Rev. 00
Klausgrete2013-08-08
Sheet
of
7
8
66015-751-60
RS485 Data Interface
Non-hazardous area
Hazardous area
Note:
1: Per circuit
2: Combined circuits
4: BAS01ATEX7005; II (1) GD [EEx ia] IIC
5: Data for CIXS3
6: Data for CAIXS2, SIWXS…-3-..-... and ISX...-..-...
7: Both channels connected on Z966
8: Only 2 RS232 connections are used on the YDI05-Z
*: Versus GND; **: Between the lines; ***: Resistively limited
YDI05-Z.. interface converter
II (2) GD [EEx ib] IIC
or
Z966 Zener barrier
4)
(in YDI01-Z) II (2) G [EEx ib] IIC or
YCO01-Y interface converter
II (2) GD [EEx ib] IIC
or
II 3 (2)GD EEx nR[ib]IIC T4
Up to 7 additional
CIXS3 or CAIXS2 /
SIWXS…-3-..-... /
ISX...-..-...
3)
Data cable:
Recommended: Sartorius cable
YCC485-X with approx.
10µH/ohm and 120pF/m (wire/wire)
up to 1000m flexibly installed.
Z966
1)
pins A/J/K/N and M
Uo 12 V*
24V**
Io 82 mA /164mA
7)
Po 0.24 W / 0.48W
7)
Co 1.41 µF*
125nF**
Lo 5.52 mH
1.38mH
6)
Lo/Ro 147 µH/
57 µH/
6)
YDI05-Z..
2) 8)
Uo 12.4 V* Ui 12.6 V*
24.8V** 25.2 V**
Io 130 mA*** Ii any
Po 400 mW Pi any
Co 1.24 µF* Ci 0
112nF**
Lo 0.4 mH* Li 0
0.4 mH**
Lo/Ro 44 µH/
* / 22 µH/ **
YCO01-Y
2)
Uo 7.2 V* Ui 12.6 V*
8.0 V** 25.2 V**
Io 207 mA*** Ii 1.5 A
Po 330 mW Pi 2.5 W
Co 13.5 µF* Ci 1 nF
8.4 µF**
Lo 0.7 mH* Li 1.6 µH
0.7 mH**
Lo/Ro 36 µH/
* / 36µH/ **
RS485 data interface
2) 6)
Ui see below Uo 5.2 V
Ii see bwlow Io 210 mA***
Pi any Po 263 mW
Ci 260 nF Co 60 µF
Li 0 mH Lo 600 µH
Lo/Ro 125 µH/
Ui ±12.4V 12.0V 7.2V
Ii 130 mA*** 164mA*** any
RS485 data interface
5) 2)
(LV4)
Uo 7.2 V
3)
Ui 12.6 V
8.2 V
4)
Io 168 mA* Ii 1.5 A*
Po 0.25 W Pi 2.5 W
Co 13 µF
3)
Ci 300 nF
3)
Co 7.6 µF
4)
Ci 100 nF
4)
Lo 0.8 mH Li 0 mH
Lo/Ro 118 µH/ohm
RS485 data interface
5) 2)
(UNICOM. LV2)
Uo 7.2 V Ui 12.6 V
Io 127 mA* Ii 1.5 A*
Po 0.273 W Pi 2.5 W
Co 11.3 µF Ci 0
Lo 2 mH Li 2 µH
Lo/Ro 118 µH/ohm
CAIXS2
II 2 G Ex ib IIC T4 Gb
II 2 D Ex ib IIIC T80°C Db
Zone 1, 2, 21, 22
Gas: IIC T4
Dust: IIIC

Table of Contents