170 Operating Instructions Combics CAIXS2
FM Certificate
FM APPROVALS PROJECT NO: 3049923
Page 9 of 10
The equipment, identified in 1.4.3, has been assessed against and found to be in
conformity with, the Essential Health and Safety Requirements (EHSR’s) of the ATEX
Directive. The record of the EHSR assessment has been retained by FM Approvals.
4 MARKING
4.1 Product intended for use in Canada shall be provided with caution and warning labels in
English and French.
4.2 A copy of label drawing 1000004644 is attached to this report. The labels contain all the
information as required by the standards.
5 REMARKS
5.1 For guidance on U.S. installations, see ANSI/ISA-RP12.06.01, Installation of Intrinsically
Safe Systems for Hazardous (Classified) Locations.
5.2 Installations shall comply with the manufacturer’s instructions.
5.3 Tampering and replacement with non-factory components may adversely affect the safe
use of the system.
5.4 The products(s) discussed in this report were certified by FM Approvals under a Type 3
Certification System as identified in ISO Guide 67.
5.5 Installations in the US shall comply with the relevant requirements of the National
Electrical Code® (ANSI/NFPA-70 (NEC®).
5.6 Installations in Canada shall comply with the relevant requirements of the Canadian
Electrical Code (CSA C22.1).
5.7 Installations in Europe shall comply with the relevant requirements of EN 60079-14 and
applicable National regulations.
5.8 FM Approvals accepts no responsibility for the compliance of the equipment identified in
1.4.3 against all applicable Directives in all applications.
6 SURVEILLANCE AUDIT
The design and manufacturing facilities at the following location(s) shall be visited on a
routine basis. The facility processes and quality control procedures in place have been
determined to be satisfactory to manufacture product identical to that tested and
Approved. A Form 797 shall be submitted to FM Approvals for requesting to
manufacture product at any additional or alternate manufacturing facilities which are not
listed below.
Design Manufacturing
Sartorius Industrial Scales GmbH & Co. KG
Sartorius Industrial Scales GmbH & Co. KG
Leinetal 2,
37120 Bovenden
Germany
Leinetal 2,
37120 Bovenden
Germany