44 Operating Instructions Combics CAIXS2
Operation
Example 1
External calibration and manual adjustment with default weights (weighing
parameters: factory settings).
(
1.) Zero the scale.
J
2.) Start calibration (e.g. when adjustment prompt flashes WP symbol).
cal.Ext. is displayed for two seconds.
You are prompted to place the required weight on the platform
(e.g. 10,000 g).
3.) Position the calibration/adjustment weight on the weighing platform.
The difference between
the weight value and the true weight of the sample
will be displayed with plus/minus signs.
Ext. calibration A printout will be generated
Targ. + 10000 g if the process is canceled
Diff. + 1 g using the ( key.
J 4.) Activate calibration/adjustment
(press the ( key to cancel).
The adjustment weight is displayed
once adjustment is finished.