Operating Instructions Combics CAIXS2 45
Operation
-------------------- A GMP-compliant printout is generated.
02/24/2013 10:15
Type CAIXS2
Ser.no. 12345678
Vers. C2 100.280810
BVers. 01-26-03 Software versions 01-26-03 to 01-26-30 can be printed.
--------------------
Ext. Calibrate
Targ. + 10000 g
Diff. + 1 g
Ext. Adjustment
Diff. + 0 g
--------------------
02/24/2013 10:15
Name:
--------------------
Example 2
External calibration and manual adjustment with freely selectable adjustment
weight (in the range 1/3 maximum load to maximum load).
(
1.) Zero the scale.
J
2.) Start calibration
(e.g. when adjustment prompt flashes WP symbol).
cal.Ext. is displayed for two seconds.
You are prompted to place the required weight on the platform (e.g. 10,000 g).
5 + )
3.) Using the number block, enter the desired CAL weight
steps and confirm with ) (in this example, 5000.0 g).
If the weight is too great or too small, an error message is displayed.