Operating Instructions Combics CAIXS2 73
Operation
Example 1: Checkweighing samples with a target weight of 1250 g and a tolerance range from
-10 g to +30 g. The tolerance values should be entered as absolute values (lower
and upper tolerance limit).
Configuration: The “Checkweighing" application is selected using the setting
INPUT / TAR.MN.MX, a printout has been set up (see “Configuration").
O t Start target value and tolerance entry using the O key.
y The target value symbol flashes at the top of the display.
t Place a sample with the target weight (in this example, 1250 g) on the
platform.
O t Save the target value.
y The minimum symbol flashes at the top of the display.
1240 t Enter lower limit value (in this example, 1240 g).
O t Save the lower limit value.
y The maximum symbol flashes at the top of the display.
1280 t Enter upper limit value (in this example, 1280 g).
O t Save the upper limit value.
y Because the sample with the target weight is still on the weighing platform,
the weight is shown on the display with the checkweighing bars. The green LED
indicates a value in the target range.
t Remove the sample with the target weight from the platform.
t The samples can now be placed on the platform and checked one after the
other.