74  Operating Instructions Combics CAIXS2
Operation
y   The LEDs next to the display indicate the results: 
yellow LED: sample too heavy 
green LED: sample in tolerance range 
red LED: sample too light.
 
p  t  Print the results. 
Note: If automatic printout of results is enabled, you do not need to press the 
p key. The results are printed automatically.
For Printout configuration: see page 96.
     Setp  +   1.250 kg    Target value
     Min   +   1.240 kg    Minimum
     Max   +   1.280 kg    Maximum
     G#    +   1.256 kg    Gross weight
     T     +   0.000 kg    Tare weight
     N     +   1.256 kg    Net weight
     Lim   +    0.48 %    Percentage of deviation from target*
     W.Diff+   0.006 kg    Absolute deviation from target
     -------------------
* When displayed in relation to target value:
If the weight is lighter than the lower limit, the display shows: LL
If the weight is heavier than the upper limit, the display shows: HH
  Example 2:  Checkweighing samples with a target weight of 1250 g and a tolerance range from 
-10 g to +30 g. The tolerance values should be entered as a relative deviation from 
the target value.
Configuration: The “Checkweighing" application is selected using the setting 
INPUT / TAR.TOL, a printout has been set up (see “Configuration").
   
  O  t  Start target value and tolerance entry using the O key.
    y  The target value symbol flashes at the top of the display.
    t   Place a sample with the target weight (in this example, 1250 g) on the 
platform.
   
  O  t  Save the target value.
y  The minimum symbol flashes at the top of the display.