Operating Instructions Combics CAIXS2 75
Operation
10 t Enter the maximum lower deviation (in this example, 10 g).
O t Save the lower limit value.
y The maximum symbol flashes at the top of the display.
30 t Enter the maximum upper deviation (in this example, 30 g).
O t Save the upper limit value.
t Proceed as described in example 1.
Example 3: Checkweighing toward zero
Y. Checkweighing samples with a target weight of
1250 g and a tolerance range from -10 g to +30 g.
Configuration: The “Checkweighing toward zero" application (TOW.ZERO) is
selected as well as the TAR.MN.MX entry, and a printout has been set up
(see “Configuration").
O t Start target value and tolerance entry using the O key.
t Place a sample with the target weight (in this example, 1250 g) ontheplatform.
O t Save the target value.
1240 t Enter lower limit value (in this example, 1240 g).
O t Save the lower limit value.
1280 t Enter upper limit value (in this example, 1280 g).