78 Operating Instructions Combics CAIXS2
Operation
By entering a percentage: The upper value of Class 1 is entered using the keypad
or by saving the value indicated. Upper limits for the other classes are defined by
entering a percentage of deviation from the upper limit of Class 1, using the
keypad.
Example: Enter 100 g as the upper limit of Class 1. Then enter 15%. When working
with 3 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g - maximum load
When working with 5 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: >minimum load - 100 g
Class 2: >100 g – 115 g
Class 3: >115 g – 130 g
Class 4: >130 g – 145 g
Class 5: >145 g - maximum load
The values entered remain valid until deleted by pressing the c key or until
overwritten by a new value. They remain saved after the scale is switched off.
Preparation t Open the APPLIC./APPLIC.2/CLASS menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 scale interval 3.6.1*
2 digit 2 scale intervals 3.6.2
5 digit 5 scale intervals 3.6.3
10 dig. 10 scale intervals 3.6.4
20 dig. 20 scale intervals 3.6.5
50 dig. 50 scale intervals 3.6.6
100 dig. 100 scale intervals 3.6.7
200 dig. 200 scale intervals 3.6.8
500 dig. 500 scale intervals 3.6.9
1000 d 1000 scale intervals 3.6.10
CTRL.SET Activate SET control output 4.3.
OUTPUT “SET" output 4.3.1*
OP.READY Ready to operate 4.3.2
OUTP.ACT port lines 4.7
OFF Off 4.7.1*
ALWAYS Always on 4.7.2
STABL. On at stability 4.7.3
QTY. Number of classes 4.8.
3 CLASS 3 classes 4.8.1*
5 CLASS 5 classes 4.8.2
INPUTParameter input 4.9.
WEIGHTS Weight values 4.9.1*
PERC.TAG Percentage values 4.9.2
PRINT Automatic printing 4.10.
MANUAL Off 4.10.1*
automat On 4.10.2
) t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.