Operating Instructions Combics CAIXS2 79
Operation
Minimum load for initialization You can set the minimum load here, i.e., the load that must be placed on the
weighing platform in order to carry out the application. If the load on platform is
too light, then this is class 0.
Setting: Applic./Applic.1/Count./Min.init menu item 3.6.
The minimum load can be set in 10 steps, from 1 to 1000 digits (see Available
parameters). The “digits" here refer to the scale intervals in the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and
1000 digits are required, you must place at least 1000 g (= 1000 intervals =
1000 digits) on the weighing platform for initialization.
Display The result of a given measurement is shown as either a weight value or a class
number.
Weight display: The current weight is shown in the measured value line and the
current class in the number display.
Display of classes: The current class is displayed in the measured value line.
Digital I|O Interface Digital I|O Interface
Control lines when working with 3 classes Control lines when working with 5 classes
Class
1
2345
Min. load
Max. load
Set
Classes 1/2
Classes 4/5
Operative
Classes 2/3/4
Class
1
23
Min. load
Max. load
Class 2
Set
Class 1
Class 3
Operative