12 Event messages
System message S
S1: Memory Error
Program and parameter memory are checked every cycle, and an error has occurred.
In the majority of cases this means that the scales is no longer ready for service.
Measures : Inform the Schenck service shop.
Parameters : O02
S5: Dongle fault
The dongle (memory for the parameters in the load cell plug) is defective or
completely missing.
Parameters : O15
Measures : Check the load cell connection
Note : INTECONT Satus is fully functional even without a dongle since
the parameters are also internally stored.
The changed parameters should be keyed in again if hardware
is replaced.
S9: Communication Host
Serial communication has been interrupted for longer than the timeout-period.
Measures : Check cable connections
Parameter : L 03
Electrical system E
E1: Power failure
The mains voltage failed or was turned off.
The flow of material may not have been registered during this time.
Measures : Acknowledge message
Parameter : O 01
E2: Tachometer Namur Error
Short circuit or break in the speed sensor cable.
The scales is no longer ready for service.
Measures : Check the cabling to the speed sensor.
Speed measuring can be temporarily turned off with parameter
B03.
Parameter : O 04
Event messages
58 BV-H2346GB / 0833 Operating Manual INTECONT® Satus