EasyManua.ls Logo

SciCan STATCLAVE G4 - Page 67

SciCan STATCLAVE G4
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67
5. Open the door. CAUTION! The rack and chamber may be hot if the unit is in standby mode.
6. Insert the temperature probe through the temperature port (TWINLOK001) on the calibration fitting
making sure the tip of the probe is well inside the chamber and not touching the rack.
7. Switch on both the temperature and pressure reference meters. Ensure the pressure meter is set to kPa
and the temperature meter is in .
8. Close the chamber door.
9. Connect one end of the pressure meter tube to your reference meter.
10. Connect the other end of the pressure meter tube to the pressure port on the calibration fitting.
STAGE 2. Starting calibration
11. From the STATCLAVE G4 home screen, access the calibration function by following these steps:
Option 1 (as provided)
Option 2 (assembly required)
Teflon washer
Knurled nut
pressure port
Temperature port
Quick connect
pressure port
Teflon washer
Temperature port

Table of Contents

Other manuals for SciCan STATCLAVE G4

Related product manuals