90
above threshold
Pressure measurement failure.
Check pressure transducer
Miscalibration.
temperature and
calculated chamber
temperature do not
Blocked exhaust / valve
Miscalibration
Valve Exhaust Chamber
filling failure
Water filling Valve bad/disconnected
Blocked tubing
Clean Water Level switch stuck OFF
filling failure
Water filling Valve bad/disconnected
Blocked tubing
External Water filling pump
bad/disconnected
IO board failure
24V power supply failure
Clean Water Level switch stuck OFF
overflow
Water reservoir exhaust blocked
Overflow switch malfunction (stuck ON)
LESS than
121°C (for 121°C cycles)
Pressure and/or temperature sensor
miscalibration.
Poor air removal during conditioning
(exhaust blockage).
Not able to generate steam or a leak in
MORE than
124°C (for 121°C cycles)
Pressure and/or temperature sensor
miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed solenoid valve).
LESS than calculated
pressure
208 kPa (for 121°C cycle)
Pressure and/or temperature sensor
miscalibration
MORE than calculated
pressure
222 kPa (for 121°C cycle)
Pressure and/or temperature sensor
miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed valve).
Poor air removal (partial exhaust