91
LESS than calculated
pressure at
205 kPa (for 121°C cycle)
Pressure and/or temperature sensor
miscalibration.
Unable to generate steam, or there is a
MORE than calculated
pressure at
225 kPa (for 121°C cycle)
Pressure sensor miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed valve).
LESS than
132°C (for 132°C cycles)
134°C (for 134°C cycles)
Pressure and/or temperature sensor
miscalibration.
Poor air removal during conditioning
(exhaust blockage).
Not able to generate steam or a leak in
MORE than
135°C (for 132°C cycles)
Pressure and/or temperature sensor
miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed solenoid valve).
LESS than calculated
pressure
291 kPa (for 132°C cycle)
309 kPa (for 134°C cycle)
Pressure and/or temperature sensor
miscalibration
MORE than calculated
pressure
309 kPa (for 132°C cycle)
327 kPa (for 134°C cycle)
Pressure and/or temperature sensor
miscalibration.Unable to depressurize
cassette (blocked exhaust, failed
valve).Poor air removal (partial exhaust
LESS than calculated
pressure at
287 kPa (for 132°C cycle)
304 kPa (for 134°C cycle)
Pressure and/or temperature sensor
miscalibration.
Unable to generate steam, or there is a
leak in the system.
MORE than calculated
pressure at
313 kPa (for 132°C cycle)
332 kPa (for 134°C cycle)
Pressure sensor miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed valve).
The time maintained by
the internal timer of the
processor didn’t match
the time maintained by
the external real-time
Steam leak