EasyManuals Logo

Servomex 1100A User Manual

Default Icon
236 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #133 background imageLoading...
Page #133 background image
5.19
5.2 DIAGNOSTIC TESTS
The function of these tests is to isolate faults at a lower level than major sub-assemblies.
5.2.1 Oxygen Transducer
The following series of tests assumes that the transducer unit is connected to a close coupled
control/interface unit in which a 01100925B amplifier board is fitted. In systems in which the
transducer is separated from the control/interface unit, refer to circuit diagrams in Figures 2.9 or
2.10 (depending upon Transducer unit fitted) for connection details.
Step Procedure Normal
Indication
Corrective
Action
1
2
3
4
5
6
Remove LED wire from terminal
TB-D-5 on amplifier board.
Measure current from TB-D-5 to
TB-D-4.
Measure current in series TB-D-5
and wire removed from TB-D-5
Refit wire into TB-D-5
Pass nitrogen as a zero gas
through the transducer at a
sample flow rate of 100 to
200ml/min
Remove photo-electric cell wiring
at TB-D-1, 2 and 3, and measure
the output from the photo-electric
cells (Wire connected to TB-D-1 is
negative polarity). Adjust
mechanical zero (see Section 4.3)
to ensure that it is possible to
cause the reading to pass through
zero. Set to 0mV +/- 5mV
50mA+/-4mA
(70mA+/-3mA
upto Serial No.
925/546)
50mA+/-4mA
(70mA+/-3mA
upto Serial No.
925/546)
Replace
Amplifier
Board
01100925B
Replace
LED in
Transducer.
See Section
5.3.2.3.
Replace
photo-cell
LED, or,
oxygen
measuring
cell.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex 1100A and is the answer not in the manual?

Servomex 1100A Specifications

General IconGeneral
BrandServomex
Model1100A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals