5.42
5.4.2 Wet Gas Sampling System 1162
5.4.2.1 Disassembly and Cleaning
See Figure 6.9.
1. Turn off gas supplies to analyser
2. Drain water out of bubbler tube by unscrewing and removing drain plug.
3. Disconnect vent/drain pipe if fitted.
4. Open door (2) by turning Dzus fastener (12) and remove door.
5. Unscrew and remove 2 X M6 slot head screws in base (4)
6. Unscrew and remove 1/4 BSP locknut (28) and collar (7) while holding bubbler base.
7. Slide glass tube and base down carefully.
8. Clean with soapy water and rinse thoroughly with clean water.
9. Replace ‘O’ rings as necessary. A complete set are in kit part no. 01162999 (Viton)
10. Reassembly is the reverse of the removal process.
Note: Do not over-tighten locknut (28).
5.4.2.2 Setting Sample Flowrate
1. Measure total sample gas flow rate out of drain / vent.
The flow meter used should have minimum back pressure.
2. Open valve slowly until a total flow rate of 200ml/min ±20ml/min is measured.
3. Adjust the sample dip leg (35) by slackening the coupling nut (33) and adjust until the flow
from the bypass dip tube just stops.
4. Re-tighten coupling nut (33).