A8.2
APPENDIX 8 SPECIFICATION
1. Functional Characteristics
Display: Digital LED display reading 0-100% oxygen with 0.01%
resolution and over range capability.
Analogue output: 0/4 - 20 mA or 20 - 4/0mA isolated, selected by the keyboard.
Maximum impedance 600 ohms.
Option of the output to be intrinsically safe "ib" when it is
connected to ground (ie non-isolated).
Spans 1, 2, 2.5, 4, 5, 10, 20, 25, 40, 50 and 100%.
The 1% span is not available with the 1100H.
2, 4, 20 and 40% spans and reverse output are not available
with software 0110066X.
Zero suppression: The zero may be suppressed in 0.01% steps to a maximum of
99.99% suppression.
Power supply: 100, 110, 117, 220, 240V ac, +/! 10% 48-62Hz 200VA.
2. Environmental Limits
Operating ambient ! 10 to +50 deg C (14 to 122 deg F) in still air. (The unit will
temperature: operate at ! 15 deg C with reduced specification).
Storage temp. ! 20 to +65 deg C (! 4 to +149 deg F).
range:
Relative humidity: 5 to 95%, non-condensing.
Atmospheric pressure 83-124kPa (12-18psia)
range:
Safety: See Section 1.2 for details of certification for use in
hazardous applications.
3. Sample Requirements (Transducer Units)
See also specification for sampling system (if supplied).
Inlet pressure 0.3kPa (50 mm wg) maximum relative to vent pressure.
to analyser: