*27 LEAK ROOM
Displays leak sensor room temperature
Normal value: (ambient + 5C) x 10
display only
*28 CELL T COR
Corrects temp-controlled flow cell at 50C
Temp difference for 50C measurement is entered (Step 0.1C)
tenkey
*29 PID P
Sets temp-controlled flow cell PID parameter
(Do not change this control parameter)
tenkey
*30 PID I
Sets temp-controlled flow cell PID parameter
(Do not change this control parameter)
tenkey
*31 PID D
Sets temp-controlled flow cell PID parameter
(Do not change this control parameter)
tenkey
*32 CUSTOMIZE
Allows the following settings at the bit level. Sets sum total of each
value. Normally setting is “2”.
2: Allows input of 0 for RESP
4: Allows 180 nm setting
8: Stops “OVER” display and continuously displays
absorbance
258: Skips wavelength check for 254 nm (Hg)
2048: Displays wavelength check results instead of SE/RE and
sends results to SCL
4096: Measures data processing time and checks REC/ITG
output (for development use)
8192: Increases digital data monotonically (for development
use)
16384: Checks digital triangular wave (1 min up to 1 mAU and 1
min down to 0 mAU) and integrity of HDLC frames (for
development use)
32768: Reverses “FUNC” logic (Normally startup -> special mode
-> If "FUNC" is pushed and a power supply switch is
turned on -> normal mode)
tenkey
*33 CLEAR ALL Initializes logs and backup memory Enter
*34 PRODUCT TYPE
Used to enter instrument type.
1: 20A 2: 20AV