Chapter 8 Troubleshooting
8.1 Using symptoms to identify the problem
Please read a see instruction manual.
8.2 Using the error display to identify the problem
No. Error Message Cause and remedial measure Error Type
1 ROM FAILURE ROM Sum error replace PCB Fatal
2 RAM FAILURE RAM Matching check error replace PCB Fatal
3 ERR GR HOME POS
Grating motor HP sensor error
loose cable, defective sensor, defective PCB
Fatal
4 ERR FIL HOME POS
Filter motor HP sensor error
loose cable, defective sensor, defective PCB
Fatal
5 ERR LAMP HOME POS
HP sensor error for lamp switching
loose cable, defective sensor, defective PCB
Fatal
6 ERR EEPROM WRITE
EEPROM writing error
defective PCB, defective device (guaranteed to
100,000 cycles)
Fatal
7 ERR LEAK DETECT
Leakage detected
Leaking liquid, thermistor contact, defective PCB
Alarm
8 NOT PROTECTED
Loss of EEPROM settings (sum error)
NMI writing error, power turned off during edit,
PCB defect, noise
Alarm
9 CHECK NO GOOD
656/254 wavelength differential greater than 1 nm.
Cell attachment, bubble intrusion, inappropriate
sample, D2 or Hg lamp defect, GM threshold defect
Alarm
10 ERR OVER CURRENT
D2 lamp anode current monitor defect
D2 lamp defect, PCB defect
Alarm
11 ERR D2 LAMP
D2 lamp will not go on even after retrying
D2 lamp defect, PCB defect
Alarm
12 ERR OVER HEAT
Overheated D2 lamp circuit radiator Open cover &
fan, D2 lamp defect, PCB defect
Alarm
13 LAMP NO RIT
Lamp does not light during ZERO & MARK
Turn on the lamp.
Warning
14 SELECT D2 SINGLE
D2 Lamp does not light in wavelength mode 2 during
WAVE CHECK
D2 malfunction, change to wavelength mode 1
Warning
15 SET ANOTHER λ
Wavelength setting defect in wavelength mode 2
Set without straddling 370/700.
Warning
16 KEY CLOSED
Key operated during KEY CLOSE mode
Press Del+CE to cancel KEY CLOSE
Warning
17 DATA NOT EXIST
Spectrum outputted without scanning
Perform scan
Warning
18 LOW SET TEMP
Cell temp set outside control range
Change temp setting
Warning
19 NO D2 LAMP
D2 lamp not installed
D2 lamp disconnected or defective, PCB defect
Alarm