(3) How to reset D/L
Using this function, all data included 5-min/30-min and parameter would be
deleted. Be sure to write down D/L parameter setting referring to D/L
parameter entry table’ in the last chapter of this manual.
Switch off power supply of D/L.
Switch on power supply of D/L and keep on pressing F5 KEY on the D/L key-
pad. Then A:> Prompt is displayed.
..
..
.. SYSTEM CONFIGRATION DISPLAY
Starting MS-DOS..
A:>
A:>444 ENTER KEY
Emergency-Mode***A:>119 ENTER KEY
Input 444 and 119 in the above screen using key-pad.
|
Press Any Key to Continue ENTER KEY
Insert new diskette for Drive C:
And press Enter when ready…ENTER KEY
Formatting 1.44Mb
Format complete
Volume Lavel (Enter for new) ? ENTER KEY
|
Format another (Y/N) ? ALT + 3
With key operation like the above picture, D/L program in A: DRIVER (embedded
in ROM) is copied in C: DRIVER and program is executed automatically.
Reset all parameters and check normal operation of D/L.