5.6.7.2 Device information
Menu: System Control Unit/Diagnosis/Device
This menu shows the version numbers of the SCU.
5.6.7.3 Cyclic trigger (CTi)
Menu: System Control Unit/Diagnosis/Cyclic Trigger
List of the next start timepoints.
Setting cyclic trigger parameters: “SCU Technical Information“ manual.
PFO (PowerFault) Fault in internal voltages Check voltages on the CAN nodes.
CONF (Config.Err) Found modules do not com-
ply with presetting
Adjust modules with default.
COM (I2C-Communication) Communication fault on node
N0
Check firm seating of I/O modules.
Global Failure Status of an analyzer Error in analyzer.
Uncertain
“U” classification in logbook, status field on operator panel (
p. 33, §5.4) lights yellow
Global Failure Status of an analyzer Error in analyzer.
Initializing Connection is being estab-
lished
----
Check
“C” classification in logbook, status field on operator panel (
p. 33, §5.4) lights yellow
Global Failure Status of an analyzer Error in analyzer.
1
If fault persists: Contact SICK Customer Service.
Have these numbers available when you have a service request concerning the
SCU.
Designation Remark
CTi Cyclic trigger name
Logbook entry Description Possible cause/ clearance
1
/System Control Unit/Diagnosis/Device
SCU
MCS100FT
Measure
xxxxxxxx
yyyyyyyy
Serial Number
Installation Package
/System Control Unit/Diagnosis/Cyclic Trigger
SCU
MCS100FT
Measure
NULLCT 1
NULLCT 2
NULLetc.